Exhaust Fan Temperature Switch (open access)

Exhaust Fan Temperature Switch

The 13000 cfm 'emergency' vent fan must be protected from over cooling which would result in a mechanical failure. Over cooling could result from a catastrophic cryogen release from the cryostat(s) or Argon Storage Dewar. In order to protect the fan, a VPT has been calibrated for -31 C to open a switch which sends a signal to allow warm gas to enter the sump by means of a motor controlled louver installed at 'sidewalk level' in the ductwork between the assembly hall and the Argon Dewar Enclosure. The bulb of the VPT is enclosed in a thermal well and will be placed in the gas stream directly above the fan. The switching unit will be mounted nearby on the wall in order to isolate it from vibrational effects. Should the fan be activated due to a cryogen release, it should not experience any problems when operating above -40 C. The switch was set and checked in a saturated calcium chloride solution cooled to -31 C by running cold gaseous Nitrogen through a copper tube coiled in a dewar. Switching temperature was measured by a thermocouple tied to the VPT bulb. The thermocouple was checked in LN2, in an ice …