A Compton scatter camera for spectral imaging of 0.5 to 3.0 MeV gamma rays (open access)

A Compton scatter camera for spectral imaging of 0.5 to 3.0 MeV gamma rays

A prototype Compton scatter camera for imaging gamma rays has been built and tested. This camera addresses unique aspects of gamma-ray imaging at nuclear industrial sites, including gamma-ray energies in the 0.5 to 3.0 MeV range and polychromatic fields. Analytic models of camera efficiency, resolution and contaminating events are developed. The response of the camera bears strong similarity to emission computed tomography devices used in nuclear medicine. A direct Fourier based algorithm is developed to reconstruct two-dimensional images of measured gamma-ray fields. Iterative ART and MLE algorithms are also investigated. The point response of the camera to gamma rays of energies from 0.5 to 2.8 MeV is measured and compared to the analytic models. The direct reconstruction algorithm is at least ten times more efficient than the iterative algorithms are also investigated. The point response of the camera to gamma rays energies from 0.5 to 2.8 MeV is measured and compared to the analytic models. The direct reconstruction algorithm is at least ten times more efficient than the iterative algorithms and produces images that are, in general, of the same quality. Measured images of several phantoms are shown. Important results include angular resolutions as low as 4.4{degrees}, reproduction of phantom …
Date: December 31, 1994