Alpha Particle Detection and the A. C. Operated Detector "A" (Howler) (open access)

Alpha Particle Detection and the A. C. Operated Detector "A" (Howler)

Abstract: This discussion concerns the operation, construction, and servicing of an alpha rate meter of medium sensitivity (30,000 alpha particles full scale per minute) and good stability. The instrument is equipped with an oscillator which increases both in frequency and loudness as the alpha activity is increased. The oscillator circuit includes an anticipator which causes a change in frequency depending directly upon the rate of change of alpha activity. Maximum sensitivity is approximately 100 millivolts full scale. Grid current varies from 1 to 5 x 10^-14 amperes. A general discussion of time constants and accuracy is included. Tables of stopping power, ionization and alpha activity are appended.
Date: January 29, 1946
Creator: Neil, Hugh G.
System: The UNT Digital Library