Livermore flow cytometer
A stable, easy to operate, flow cytometer was built for routine use by research personnel in a biology laboratory. A 5 watt argon-ion laser excites the fluorescent dyes in cells flowing through a quartz flow chamber. A f/0.95 TV camera lens collects the fluorescent and scattered light and directs it through color separation filters and a beam splitter onto two photomultiplier tubes. Two 3-axis positioners are used to align the focused laser beam and the flow chamber with the light collection optics. All of the components, except the laser, are mounted on a rigid aluminum plate. Standard electronics modules are used. A coefficient of variation of about 1% can be obtained with fluorescent microspheres.
Date:
November 19, 1981
Creator:
Peters, D.C.; Dean, P.N. & Pinkel, D.
Object Type:
Report
System:
The UNT Digital Library
Proceedings of the twenty-third WANTO meeting: a workshop to discuss policies and practices used to assure quality NDE operations
Ten papers were presented at the meeting. A separate abstract was prepared for each paper. (LCL)
Date:
November 19, 1981
Creator:
Baxter, G.R. (comp.)
Object Type:
Article