The Use and Calibration of Scintillation Counter--Model Y (open access)

The Use and Calibration of Scintillation Counter--Model Y

Several improvements have been made in the gamma scintillation counter (GSC). The Model V gamma scintillation counter uses a canned thallium activated sodium iodide crystal as a detector. Although the electronic components remain unchanged, a modification of the sample support has been made to improve reproducibility of geometry. To assure comparable results between various counters, they must be operated at the same energy threshold and counting yield. Methods have been developed to assure operation at a preferred energy threshold of 0.1 Mev and an arbitrarily fixed counting yield of 3.36% for Cs-Ba-137. Partial pulse height discrimination against U 237 is also accomplished.
Date: August 1, 1953
Creator: Brauer, F. P. & Leboeuf, M. B.
System: The UNT Digital Library